Two Human Hydroxyapatite-Coated Dental Implants Retrieved After a 14-Year Loading Period: A Histologic and Histomorphometric Case Report

2007 
Background: Controversy over the long-term clinical effectiveness of hydroxyapatite (HA)-coated dental implants still persists, despite numerous clinical studies documenting high survival rates. Concerns about the degradation of the coating over the years have been raised; it has been speculated that resorption of the HA could produce a space between the implant and the bone with a resultant mechanical instability.Methods: Two HA-coated implants were retrieved due to a fracture of the abutment screws after a loading period of 14 years and were treated to obtain thin ground sections for histologic evaluation.Results: At low-power magnification, it was possible to observe that the HA coating was in contact with mature bone. No gaps or connective fibrous tissue was found at the implant–bone interface. No epithelial downgrowth was present. No acute or chronic inflammatory cell infiltrate was present at the implant–bone interface. No foreign body reaction was present in the peri-implant tissues. Some osteocyte...