Hypoglycemia Detection and Carbohydrate Suggestion in an Artificial Pancreas

2016 
Fear of hypoglycemia is a major concern for many patients with type 1 diabetes and affects patient decisions for use of an artificial pancreas system. We propose an alternative way for prevention of hypoglycemia by issuing predictive hypoglycemia alarms and encouraging patients to consume carbohydrates in a timely manner. The algorithm has been tested on 6 subjects (3 males and 3 females, age 24.2 ± 4.5 years, weight 79.2 ± 16.2 kg, height 172.7 ± 9.4 cm, HbA1C 7.3 ± 0.48%, duration of diabetes 209.2 ± 87.9 months) over 3-day closed-loop clinical experiments as part of a multivariable artificial pancreas control system. Over 6 three-day clinical experiments, there were only 5 real hypoglycemia episodes, of which only 1 hypoglycemia episode occurred due to being missed by the proposed algorithm. The average hypoglycemia alarms per day and per subject was 3. Average glucose value when the first alarms were triggered was recorded to be 117 ± 30.6 mg/dl. Average carbohydrate consumption per alarm was 14 ± 7.8...