CORTICAL BONE SPHERICAL INDENTATION MODULUS CORRELATES WITH THE ELASTIC MODULUS FROM COMPRESSION TESTING AND CT-MEASURED POROSITY

2019 
With information about a patient's bone mechanical properties orthopaedic operations could be optimised to reduce intra- and post-operative complications. However, there is currently no reliable method of measuring a patient's bone mechanical properties in vivo. We have previously investigated microindentation, using a 1.5mm diameter spherical indenter tip, and found no correlation between these measurements and compression testing measurements. We hypothesised that by using a larger diameter indenter tip we would closer match bone millimetre-scale mechanical properties. 20 bone samples were taken from 20 patients undergoing hip replacement surgery. The samples were machined from the femoral neck calcar cortical bone into 6×3×3mm parallelepiped specimens, aligned with the osteons along the long axis. The samples were micro-computed tomography (CT) scanned to calculate porosity. Microindentation was performed using a 6mm diameter, sapphire, spherical indenter tip. 12 indentations were performed in a grid a...