Simultaneous Assessment of Fractional and Coronary Flow Reserves in Cardiac Transplant Recipients

2003 
Background— The utility of measuring fractional flow reserve (FFR) to assess cardiac transplant arteriopathy has not been evaluated. Measuring coronary flow reserve (CFR) as well as FFR could add information about the microcirculation, but until recently, this has required two coronary wires. We evaluated a new method for simultaneously measuring FFR and CFR with a single wire to investigate transplant arteriopathy. Methods and Results— In 53 cases of asymptomatic cardiac transplant recipients without angiographically significant coronary disease, FFR and thermodilution-derived CFR (CFRthermo) were measured simultaneously with the same coronary pressure wire in the left anterior descending artery and compared with volumetric intravascular ultrasound (IVUS) imaging. The average FFR was 0.88±0.07; in 75% of cases, the FFR was less than the normal threshold of 0.94; and in 15% of cases, the FFR was ≤0.80, the upper boundary of the gray zone of the ischemic threshold. There was a significant inverse correlati...