Adjustable- Versus Fixed-Loop Devices for Femoral Fixation in ACL Reconstruction: An In Vitro Full-Construct Biomechanical Study of Surgical Technique–Based Tibial Fixation and Graft Preparation:

2018 
Background:Femoral suspensory fixation for anterior cruciate ligament (ACL) reconstruction has evolved from fixed- to adjustable-loop devices. However, there are still controversies regarding undesired lengthening of adjustable-loop devices.Hypothesis:Adjustable-loop fixation will achieve similar elongation to that of fixed-loop devices, and intraoperative preconditioning will reduce initial elongation for adjustable-loop constructs.Study Design:Controlled laboratory study.Methods:Three adjustable-loop devices (GraftMax, TightRope, and Ultrabutton) and 2 fixed-loop devices (Endobutton and RetroButton) were used in an intraoperative surgical technique workflow according to an in vitro model with porcine bone and bovine tendons (8 specimens per device; N = 40 constructs tested). Each construct underwent 1000 cycles of position- and force-controlled dynamic loading, whereby a total elongation threshold of 3 mm was defined as clinical failure. Constructs were finally pulled to failure at 50 mm/min.Results:The...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    20
    Citations
    NaN
    KQI
    []