Development of cryogenic mechanisms for the VLT/ERIS instrument

2018 
We present the design and measured performance of the Aperture Wheel and the Pupil and Filter Wheel mechanisms for the NIX camera of the VLT/ERIS instrument. Both mechanisms were developed for high opto-mechanical precision and stability while operating at 70 K. We summarise the design constraints and considerations. Further, we have developed a dedicated cryo-test facility to allow measuring the position repeatability under nominal operational conditions. We demonstrate that the wheel mechanisms perform as designed and provide the measurement methodology and results of the opto-mechanical tolerances.