Does the cleaning technique influence the durability of the <9F flexible ureteroscope?

2001 
Background and Purpose: The fragility of the <9F flexible ureteroscope limits its availability to general urology practice. The purpose of this study was to determine whether the technique used to clean the flexible ureteroscope or the number of persons handling the instrument during the cleaning process influenced endoscope breakage or deterioration during regular endourologic use. Patients and Methods: A new Olympus URF/P3 flexible 7.5F ureteroscope was used for each of two 30-day study periods during which a single surgeon used the endoscope for a variety of upper urinary tract procedures. During the first 30-day period (Group 1), the endoscope was leak-proof-pressure tested and cleaned by the endourology support team using the Steris 20 (peroxyacetic acid 35%) technique. During the second 30-day period (Group 2), the endoscope was leak-proof tested and cleaned only by the surgeon using the Cidex (glutaraldehyde 2.4%) technique. A record was kept for each ureteroscopic case to document the patient posi...