Protected Carotid Stenting in High‐Risk Patients: Results of the SpideRX Arm of the Carotid Revascularization with ev3 Arterial Technology Evolution Trial

2010 
Purpose: A prospective nonrandomized multicenter registry of 160 patients with severe carotid stenosis and high-risk features for carotid endarterectomy was conducted during the 3-month period from March to May 2005. Methods: Carotid artery stenting (CAS) was performed with the SpideRX™ Embolic Protection System (ev3, Inc., Plymouth, MN, USA) as part of an investigational device exemption from the Food and Drug Administration. Results: The primary end-point of major adverse cardiac and cerebrovascular events at 30 days after CAS was observed in nine patients (5.6%), including death in four patients (2.5%), nonfatal stroke in five patients (3.1%), and nonfatal myocardial infarction in one patient (0.6%). A secondary end-point of technical success (defined as successful deployment of all devices, filter retrieval, and final diameter stenosis <50%) was achieved in 156 of 160 patients (97.5%). The only independent predictor of death or stroke at 30 days was baseline stenosis severity (P < 0.05). Conclusion: CAS with distal embolic protection using the SpideRX™ Embolic Protection System is a reasonable alternative for revascularization of some high-risk patients with severe carotid stenosis. (J Interven Cardiol 2010;23:491–498)