Amino Acids Profiles in Biological Media
2010
An accurate analytical method was developed to determine amino acids in some biological specimens by GC/MS technique. Stable isotopes provide useful tools for a variety of studies, offering ideal internal standards in quantitative information. Isotopic dilution gas chromatography—mass spectrometry (ID‐GC/MS) is the techniques used for quantitative analysis of compounds labeled with stable isotopes. A Trace DSQ Thermo Finnigan quadrupole mass spectrometer coupled with a Trace GC was used. Amino acids were separated on a Rtx‐5 MS capillary column, 30 m×0.25 mm, 0.25 μm film thickness, using a temperature program from 50 °C, 1 min, 6 °C/min at 100 °C, 4 °C/min at 200 °C, 20 °C/min at 300 °C, (3 min). The transfer line temperature was 250 °C, the injector temperature 200 °C and ion source temperature 250 °C; splitter: 10:1. Electron energy was 70 eV and emission current, 100 μA. The amino acids were purified on a Dowex 50W‐W8 exchange resin and were derivatized in a procedure following two steps to obtain tri...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
2
Citations
NaN
KQI