Salt-bridge-supported bilayer lipid membrane biosensor for determination of anticancer drug cyclophosphamide

2001 
A novel biosensor for assaying anticancer drug cyclophosphamide was constructed with salt-bridge supported bilayer lipid membrane modified with tetraphenylborate- cyclophosphamide complex. The modification was achieved by the introduction of the complex into the membrane forming solution. The biosensor show a linear response to the drug over the concentration range 8.96 X 10 -6 mol L -1 . The effects of coexistent substances and pH on assay were evaluated. The results show that the distinguish merits of this kind of biosensor is the excellently biological compatibility and no need of mediator for ions exchange. It also shows good selectivity and sensitivity for cyclophosphamide assay.