Cost Analysis of Flexible Ureteroscope Repairs: Evaluation of 655 Procedures in a Community-Based Practice

2016 
Abstract Introduction: The frequency of flexible ureteroscopy has increased with the introduction of improved instrumentation. Ureteroscopes allow increased endoscopic access to the ureter and kidney. However, maintenance and repair of scopes may increase the total procedure expense. Methods: In 3 years (8/2011–7/2014), 655 flexible ureteroscopies were performed at a single-specialty, urology, ambulatory surgery center. Procedures were performed by 26 board-certified urologists using four Olympus URF P5 flexible ureteroscopes. The instruments were handled by a single team and sterilized through the STERIS System E1. Repairs were performed by the manufacturer on an as needed basis. Patient records were reviewed to determine the preoperative diagnosis, operative time, location and size of the stone, and use of laser or ureteral sheath. The occurrence, nature of flexible ureteroscope damage, and cost of repairs were evaluated. Results: Of the ureteroscopies performed, 78% was for the treatment of calculi (50...