An Example Crossover Experiment for Testing New Vicarious Calibration Techniques for Satellite Ocean Color Radiometry

2010 
Abstract Vicarious calibration of ocean color satellites involves the use of accurate surface measurements of water-leaving radiance to update and improve the system calibration of ocean color satellite sensors. An experiment was performed to compare a free-fall technique with the established Marine Optical Buoy (MOBY) measurement. It was found in the laboratory that the radiance and irradiance instruments compared well within their estimated uncertainties for various spectral sources. The spectrally averaged differences between the National Institute of Standards and Technology (NIST) values for the sources and the instruments were <2.5% for the radiance sensors and <1.5% for the irradiance sensors. In the field, the sensors measuring the above-surface downwelling irradiance performed nearly as well as they had in the laboratory, with an average difference of <2%. While the water-leaving radiance Lw calculated from each instrument agreed in almost all cases within the combined instrument uncertainties (a...