Fibreoptic intubation. Influence of tracheal tube tip design.

1993 
A randomised study was carried out in 60 patients to assess the influence of tracheal tube tip design on the ease of railroading a tracheal tube during fibreoptic intubation. A new design of tracheal tube with a tapered tip, without a bevel, was compared with a tracheal tube of standard design. The new design was found to be greatly superior in both oro- and nasotracheal fibreoptic intubation, when compared with the traditional tracheal tube (p<0.001). The shape of the tip of a tracheal tube is an important determinant of the ease of railroading the tube over an inserted fibrescope