Low-dose CT screening using hybrid iterative reconstruction: confidence ratings of diagnoses of simulated lesions other than lung cancer

2015 
Objective:To evaluate the confidence ratings of diagnoses of simulated lesions other than lung cancer on low-dose screening CT with hybrid iterative reconstruction (IR).Methods:Simulated lesions (emphysema, mediastinal masses and interstitial pneumonia) in a chest phantom were scanned by a 320-row area detector CT. The scans were performed by 64-row and 160-row helical scans at various dose levels and were reconstructed by filtered back projection (FBP) and IR. Emphysema, honeycombing and reticular opacity were visually scored on a four-point scale by six thoracic radiologists. The ground-glass opacity as a percentage of total lung volume (%GGO), CT value and contrast-to-noise ratio (CNR) of mediastinal masses were calculated. These scores and values were compared between FBP and IR. Wilcoxon's signed-rank test was used (p < 0.05). Interobserver agreements were evaluated by κ statistics.Results:There were no significant differences in visual assessment. Interobserver agreement was almost perfect. CT value...