NUMERICAL ANALYSIS OF BALLOON EXPANDABLE STENT DEPLOYMENT INSIDE A PATIENT-SPECIFIC STENOTIC CORONARY ARTERY TO INVESTIGATE THE INSTANT MECHANICAL BEHAVIORS

2017 
The instant mechanical behaviors of stenotic coronary artery and deployed stents have significant impacts on percutaneous coronary intervention prognosis. However, they could not be obtained directly from the current examination techniques, which are commonly used in clinical practice. Thus, we intend to investigate the instantaneous mechanical behaviors of deployed stent and artery through virtually stenting technology based on a real clinical case in assessment of geometric and biomechanical characteristics. Method: Finite element analysis models, including rigid guide catheter, six-folded balloon with conical tip, crimped and bended stent, stenotic coronary artery with soft plaques, were simulated through virtual mechanical expansion and recoil procedure. The morphology changes of coronary lumen, strain and stress distribution of involved components at different stages and apposition of stent struts were analyzed. Results: Lumen in the stenotic region restored patency obviously at maximum expansion and...