A new endo aortic occlusion balloon for limited access cardiac surgery: development and clinical evaluation.

2001 
In recent years, minimally invasive cardiac surgery (MICS), or limited access cardiac surgery, has been presented as a promising operative procedure. We developed a new balloon device that is inserted directly into the ascending aorta to stop the heart during limited access cardiac surgery. The balloon has a three lumen structure: balloon lumen port, cardioplegia/vent lumen port, and aortic root lumen port. This direct EAC balloon catheter, designed to be inserted directly into the ascending aorta, is different from the Heartport system. The Heartport EAC balloon catheter is inserted into the aorta via an artery in the lower limb, making lower limb arterial disease a key concern. Our Direct Endo Aortic Clamp (EAC) balloon overcomes this problem. The device was clinically used in seven cardiac cases. All patients were discharged within 5 postoperative days, confirming the utility of the device.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader