Emergency HeartWare Ventricular Assist Device (HVAD) exchange due to pump thrombosis using minimally invasive technique

2017 
: Left ventricular assist device (LVAD) thrombosis remains a dreadful complication of mechanical circulatory support, with an incidence of 8-12% depending on the pump type and patient's comorbidities. Fibrinolysis may be considered early in pump thrombosis, but when contraindicated a pump exchange remains the only alternative. This short report documents an emergency LVAD exchange in a 55-year-old man who underwent LVAD (HeartWare Inc) implantation in 2013 as a bridge to transplantation. Four months after the initial surgery, he suffered from a hemorrhagic stroke despite properly managed anticoagulation. On February 17th, 2017 he was re-admitted with LVAD pump thrombosis. As fibrinolysis was contraindicated, an emergency pump exchange was performed via a limited thoracic incision in order to minimize surgical trauma, reduce intraoperative complications and facilitate immediate post-operative recovery. This report documents the very first LVAD pump exchange as well as the first one performed via a minimally invasive approach in Poland.