Evaluation of a Novel Immunochromatographic Assay for Detection of Human Adenoviruses from Throat Swab Samples Compared with Real-time PCR

2018 
BACKGROUND: The aim of this study was to determine the sensitivity and specificity of a novel immunochromatographic assay (ICA) kit, ALSONIC® Adeno (Alfresa Pharma Co., Osaka, Japan), for the detection of human adenovirus (HAdV) from throat swab samples based on the results of real-time PCR. The incubation time required for the novel assay kit (5 minutes) is shorter than that required for other ICA kits that are available in Japan. METHODS: Throat swab samples were taken from 151 patients aged 6 months to 15 years who were suspected of having respiratory tract infections caused by HAdV. RESULTS: The sensitivity and specificity of the ICA for detection of HAdV were 92.2% (83/90) and 95.1% (58/61), respectively, and the assay showed positive and negative predictive values of 96.5% (83/86) and 89.2% (58/65), respectively. CONCLUSIONS: ALSONIC® Adeno is suitable as a diagnostic tool in the acute phase of HAdV infection.