Evaluation and Specifications for In-Line Uranium Separations Using Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES) Detection for Trace Elemental Analysis:

2019 
Automated introduction platforms integrated with inductively coupled plasma – optical emission spectroscopy (ICP-OES) systems are continuously being improved. Expanding on the introduction systems, a newly developed automated ion chromatography system was explored for performing rapid in-line separations coupled to ICP-OES for the detection of trace elements in uranium. Trace elements are separated from a uranium material and the analytes are directed into the ICP-OES for subsequent detection. Detection parameters such as exposure time frequency, wavelength selection, and settling times were explored to gain insight on optimal detection schemes for in-line trace elemental analysis. The methodology was applied in the analysis of a uranium oxide (U3O8) certified reference material, CRM-124. It was found here that the sensitivity and uncertainty of the technique are greatly affected by how the ICP-OES is employed to collect data. Overall it was determined that faster exposure replicates can provide greater p...