Abstract Purpose This report describes complications using the vertical expandable prosthetic titanium rib (VEPTR) for thoracic insufficiency syndrome (TIS) at a single center. Methods This is a prospective cohort evaluating 65 patients with rib–rib and rib–spine VEPTR devices for TIS placed between 10/2001 and 11/2014, for children with spinal or chest wall deformity. Patients were classified using the early onset scoliosis classification system (C-EOS). Results 65 patients are available for follow up. 23 congenital scoliosis, 12 neuromuscular, 14 syndromic, 2 idiopathic and 14 not classifiable by the C-EOS system including 11 chest wall reconstructions. Average age at implantation was 6.9years (range 1.3–24.8) with average follow up 6.9years (range 0.4–14.8). 22 patients had 37 complications. Those classifiable by C-EOS had complications in the normo- and hyperkyphotic groups. Implant erosion and infection were most common. The majority of complications required one additional unplanned surgery for resolution. Two complications required abandonment of a growth-friendly strategy. Conclusions Use of VEPTR for TIS is associated with significant and frequent complications. C-EOS suggests that complications are more likely in those with normal or hyperkyphotic curves. Most complications are managed with one unplanned surgery. VEPTR is usually salvaged and abandonment of a growth-friendly strategy is unusual.