Efficacy of Diabeloop closed‐loop to improve glycaemic control in patients with type 1 diabetes exposed to copious dinners or to sustained physical exercise

2019 
AIMS: To compare closed-loop (CL) and open-loop (OL) systems for glycaemic control in patients with type 1 diabetes (T1D) exposed to real-life challenging situations (gastronomic dinners or sustained physical exercise). METHODS: Thirty-eight adult patients with T1D were included in a three-armed randomized pilot trial (Diabeloop WP6.2 trial) comparing glucose control using a CL system with use of an OL device during two crossover 72-hour periods in one of the three following situations: large (gastronomic) dinners; sustained and repeated bouts of physical exercise (with uncontrolled food intake); or control (rest conditions). Outcomes included time in spent in the glucose ranges of 4.4-7.8 mmol/L and 3.9-10.0 mmol/L, and time in hypo- and hyperglycaemia. RESULTS: Time spent overnight in the tight range of 4.4 to 7.8 mmol/L was longer with CL (mean values: 63.2% vs 40.9% with OL; P 10.0 mmol/L) compared to those using an OL system (17.9% vs 31.9%; P
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    42
    References
    13
    Citations
    NaN
    KQI
    []