A Comparison of Radiostereometric Analysis and Computed Tomography for the Assessment of Lumbar Spinal Fusion in a Sheep Model

2013 
Study Design  Prospective animal study. Objective  The aim of this animal study is to evaluate the accuracy of radiostereometric analysis (RSA) compared with computed tomographic (CT) scan in the assessment of spinal fusion after anterior lumbar interbody fusion (ALIF) using histology as a gold standard. Methods  Three non-adjacent ALIFs (L1–L2, L3–L4, and L5–L6) were performed in nine sheep. The sheep were divided into three groups of three sheep. All the animals were humanely killed immediately after having the last scheduled RSA. The lumbar spine was removed and in vitro fine cut CT and histopathology were performed. Results  Using histological assessment as the gold standard for assessing fusion, RSA demonstrated better results (100% sensitivity and 66.7% specificity; positive predictive value [PPV] = 27.3%, negative predictive value [NPV] =100.0%) compared with CT (66.7% sensitivity and 60.0% specificity [PPV = 16.7%, NPV = 93.8%]). Conclusions  RSA demonstrated higher sensitivity and specificity when compared with CT. Furthermore, RSA has the advantage of much lower radiation exposure compared with fine cut CT. Further studies are required to see if RSA remains superior to CT scan for the assessment spinal fusion in the clinical setting.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    6
    Citations
    NaN