Comparison between PRIMO and EGSnrc Monte Carlo models of the Varian True Beam linear accelerator

2020 
Abstract The study objectives were to simulate a 6 MV flattening filter-free (FFF) photon beam of a TrueBeam Varian LINAC, using the PRIMO code. The depth-dose profiles for various jaw open fields and cross-beam profiles for various depths inside water phantoms were determined using PRIMO simulations and were compared with the EGSnrc simulation results. A 93% match between both codes has generally been obtained, with a systematic variation in agreement between the two models that increases with field size. The closest agreement, to within 94%, was obtained for a 6 × 6 cm2 jaw open field at 5.0 cm and 10.0 cm depths, better at 10- than at 5.0 cm depth. The effects of how scatter is modelled has been shown to result in the differences.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    0
    Citations
    NaN
    KQI
    []