Accelerated MRI of the knee. Quality and efficiency of compressed sensing

2020 
Abstract Purpose To evaluate potential clinical acceleration factors of Compressed SENSE (CS) 1 in direct comparison with SENSE for fat saturated (fs) 2 , proton density-weighted (PD) 3 2D and 3D sequences of the knee. Method Twenty healthy volunteers were scanned with a 3 T scanner, all receiving a standard, fs 2D PD, three CS (CS 2, CS 3, CS 5) as well as time-equivalent SENSE accelerations (S 2, S 3, S 5). The fs 3D PD sequence was acquired with four CS (CS 6, CS 8, CS 10, CS 15) and equivalent SENSE (S 5.72, S 7.69, S 9.57, S 14) factors. Three independent readers rated the images. Signal-to-noise, contrast-to-noise, root-mean-square error and structural similarity index were analyzed for objective evaluation. Results Scan time decreased with increasing CS factor (2D CS 2: 145 s, 2D CS 3: 95 s, 2D CS 5: 57 s, 3D CS 6: 293 s, 3D CS 8: 220 s, 3D CS 10: 176 s, 3D CS 15: 119 s). The 2D standard sequence was rated best for diagnostic certainty and overall image impression with an average of 4.97 ± 0.10 and 4.80 ± 0.24 (all p  Conclusion Compressed SENSE can significantly decrease (34.39 % for 2D CS 2 and 54.17 % for 3D CS 10) scan time in knee imaging with unchanged diagnostic certainty and overall image impression compared to the clinical reference.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    2
    Citations
    NaN
    KQI
    []