ESPRESSO: A novel device for laser-assisted surgery of the anterior eye segment

2016 
AbstractIntroduction: Merging robotics with laser eye surgery could enhance precision, repeatability and automation. During some eye laser procedures the patient is awake, thus eye stabilization is desired to avoid movements that could affect the treatment. Material and methods: The ESPRESSO platform has a two-stage actuation system to position a stabilization tool on the eye, a proximity sensing unit to monitor the stabilization tool position, and a sensing unit to monitor the pressure exerted on the eye. The platform is tested in-vitro and ex-vivo with clinicians. A maximum pressure to be exerted on the eye is defined with expert ophthalmic surgeons to be 22 mmHg: physiological intraocular pressure (IOP) range is 10-21 mmHg. This pressure corresponds to a force of 0.3 N. Results: The necessary contact force to have eye fixation (according to the clinicians' feedback) is evaluated: maximum values resulted always below 0.3 N. A maximum IOP increase of 4.67 mmHg is observed, that is a slight variation with...