A C OMPARISON OF B IPHASIC AND M ONOPHASIC S HOCKS FOR E XTERNAL D EFIBRILLATION

2000 
Background and Objective. The ability of a shock to defibrillate the heart depends on its waveform and energy. Past studies of biphasic truncated exponential (BTE) shocks for external defibrillation focused on low energy levels. This prospective, randomized, double-blind clinical trial compared the first-shock efficacies of 200-joule (J) BTE, 130-J BTE, and 200-J monophasic damped sine wave shocks. Methods. Ventricular fibrillation (VF) was induced in 115 patients during evaluation of implantable cardioverter-defibrillator function and 39 patients during electrophysiologic evaluation of ventricular arrhythmias. After 19 ± 10 seconds of VF, a randomized transthoracic shock was administered. Mean first-shock success rates of the three groups were compared using a “Tukey-like” statistical test, adjusting for multiple comparisons. Blood pressures and arterial oxygen saturations were measured before VF induction and 30, 90, and 150 seconds after successful defibrillation. Results. First-shock success rates wer...
    • Correction
    • Cite
    • Save
    • Machine Reading By IdeaReader
    12
    References
    58
    Citations
    NaN
    KQI
    []