High-resolution Compton-suppressed CZT and LaCl/sub 3/ detectors for fission products identification

2005 
Room-temperature semiconductor CdZnTe (CZT) detectors are currently limited to total detector volumes of 1-2 cm/sup 3/, which is dictated by poor charge transport characteristics. Because of this size limitation, one of the problems in accurately determining isotope identification is the enormous background from Compton scattering events. Eliminating this background will not only increase the sensitivity and accuracy of measurements, but will also help to resolve peaks buried under the background and peaks in close vicinity to others. We are currently developing a fission products detection system based on Compton-suppressed CZT and LaCl/sub 3/ (Ce) detectors. In this application, the detection system is required to operate in a high radiation field. Therefore, a small 10 mm /spl times/10 mm/spl times/5 mm CZT and 13mm/spl times/15mm LaCl/sub 3/ detector are placed inside the center of a well-shielded 76 mm by 76 mm long NaI detector. So far, we have been able to successfully reduce Compton background by a factor of 3.7 to 4.0 for a /sup 137/Cs spectrum. In this work, we will discuss the performance of this detection system using both CZT and LaCl/sub 3/ detectors. The results are compared with MCNP calculations.