IL8 and IL16 levels indicate serum and plasma quality

2018 
BACKGROUND: Longer pre-centrifugation times alter the quality of serum and plasma samples. Markers for such delays in sample processing and hence for the sample quality, have been identified. METHODS: Twenty cytokines in serum, EDTA plasma and citrate plasma samples were screened for changes in concentration induced by extended blood pre-centrifugation delays at room temperature. The two cytokines that showed the largest changes were further validated for their "diagnostic performance" in identifying serum or plasma samples with extended pre-centrifugation times. RESULTS: In this study, using R&D Systems ELISA kits, EDTA plasma samples and serum samples with a pre-centrifugation delay longer than 24 h had an IL16 concentration higher than 313 pg/mL, and an IL8 concentration higher than 125 pg/mL, respectively. EDTA plasma samples with a pre-centrifugation delay longer than 48 h had an IL16 concentration higher than 897 pg/mL, citrate plasma samples had an IL8 concentration higher than 21.5 pg/mL and serum samples had an IL8 concentration higher than 528 pg/mL. CONCLUSIONS: These robust and accurate tools, based on simple and commercially available ELISA assays can greatly facilitate qualification of serum and plasma legacy collections with undocumented pre-analytics.