An iterative method for the deconvolution of microcalorimeter thermograms

1987 
Abstract An iterative method for the deconvolution of microcalorimetry thermograms suitable for small digital computers is presented. The method employs a measured impulse response function directly as the deconvolution kernel, thus explicit system simulation is not required. Data are presented showing the performance of the method and the exchange of signal-to-noise ratio for time resolution that is made when deconvolution techniques are employed. An improvement in the system time resolution of fifty times is demonstrated with measured data.