Comparative study of large samples (2" × 2") plastic scintillators and EJ309 liquid with pulse shape discrimination (PSD) capabilities

2014 
In the paper we reported on the scintillation properties and pulse shape discrimination (PSD) performance of new plastic scintillators. The samples with dimension of 2 inches × 2 inches were tested: EJ299-34, EJ299-34G, EJ299-33 and EJ299-33G. They are the first commercially available plastics with neutron/gamma discrimination properties. The paper covers the measurements of emission spectra, photoelectron yield, analysis of the light pulse shapes originating from events related to gamma-rays and fast neutrons as well as neutron/gamma discrimination. The tested plastics are characterized by a photoelectron yield on a level of approximately 1600-2500 phe/MeV, depending on the sample. The highest value, measured for EJ299-34, is similar to the number of photoelectrons measured for EJ309 (2600 phe/MeV). The figure of merit (FOM) calculated for narrow energy cuts — indicating the PSD performance — showed that the PSD capabilities of the plastics are significantly lower than of EJ309. These scintillators are still under development in order to optimize the composition and manufacturing procedures. At this time the results obtained with the new plastics suggest their possible use as an alternative for liquid scintillators, especially if we consider their inflammability and non-toxicity.