Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology

2018 
Purpose To evaluate the correlation between anterior axial curvature difference maps following corneal cross-linking (CXL) for progressive keratoconus obtained from Scheimpflug-based tomography and Placido-based topography. Design Between-devices reliability analysis of randomized clinical trial data. Methods Corneal imaging was collected at a single-center institution preoperatively and at 3, 6, and 12 months postoperatively using Scheimpflug-based tomography (Pentacam; Oculus Inc, Lynnwood, Washington, USA) and scanning-slit, Placido-based topography (Orbscan II; Bausch & Lomb, Rochester, New York, USA) in patients with progressive keratoconus receiving standard protocol CXL (3 mW/cm 2 for 30 minutes). Regularization index (RI), absolute maximum keratometry (K Max), and change in K Max (ΔK Max) were compared between the 2 devices at each time point. Results Fifty-one eyes from 36 patients were evaluated at all time points. Values were significantly different at all time points (56.01 ± 5.3 diopters [D] Scheimpflug vs 55.04 ± 5.1 D scanning-slit preoperatively [ P  = .003]; 54.58 ± 5.3 D Scheimpflug vs 53.12 ± 4.9 D scanning-slit at 12 months [ P Conclusions Measurements using Scheimpflug and scanning-slit Placido-based technology are correlated but not interchangeable. Both devices appear reasonable for separately monitoring the cornea's response to CXL; however, caution should be used when comparing results obtained with one measuring technology to the other.