Exact resolution function for double-disk chopper neutron time-of-flight spectrometers : Application to reflectivity
2020
The exact resolution function in transfer vector for the reflectometer HERM\`ES at Laboratoire L\'eon Brillouin is calculated as an example of neutron time-of-fligh spectrometer with double-disk chopper. Calculation accounts for wavelength distribution of the incident beam, tilt of the chopper axis, collimation and gravity, without approximation of gaussian distributions or independence of these different contributions. Numerical implementation is provided that matches the sections of the paper. We show that data fitting using this exact resolution function allows us to reach much better results than its usual approximation by a gaussian profile.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
9
References
2
Citations
NaN
KQI