Optimization of Measuring Cardiac Output by Both Hands Electrode Impedance Method
2011
In this study, a new method that can estimate ICG data from a subject`s both hands to measure Cardiac Output under convenient sensor environment. With this aim, a grip-type electrode was implemented to measure ICG. To improve the accuracy of measurement, the regression equation was extracted using multiple bio-parameters and our result was compared with the thoracic ICG equipment(Physio Flow, PF104D, Manatec Biomedical, France), which is being used in clinics. The subjects consist of 26 men and 4 women(age in ). They are no cardiac disease. Parameters available for regression model were used gender, BMI, MBP, LVET, dZ/dt(max), distance between the measured electrodes, body impedance, and PTT. As a result of analyzing the ICG measurement value on thorax and both hands, the correlation with stroke volume, heart rate, and cardiac output was