Performance of a low voltage ion trap

2011 
Abstract Reduced power operation of a mass analyzer with minimum loss of spectral resolution and mass range is desirable in portable instruments. Miniaturizing quadrupole-based ion traps can be especially worthwhile since the RF amplitude necessary for mass analysis scales with the square of the analyzer dimensions. The performance of a miniature, stainless steel, rectilinear ion trap (RIT) with dimensions of 1.66 mm × 1.43 mm ( x 0 and y 0 respectively) is characterized by sampling externally generated ions and performing mass analysis without benefit of differential pumping to simulate conditions in a miniature system. This system is capable of detecting analyte ions of up to m / z 1250 at operating voltages of 610 V 0-p (drive frequency of 1.105 MHz), and with spectral resolution on the order of 2 Th (FWHM) across the entire mass range. The ability to acquire structural information through tandem MS is also demonstrated.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    32
    References
    12
    Citations
    NaN
    KQI
    []