Method evaluation of pepsinogen I/II assay based on chemiluminescent immunoassays and comparison with other test methods.

2016 
Abstract Background Serum pepsinogen (PG) I and the PG I/PG II ratio have been used for atrophic gastritis (AG) diagnosis for decades. Low levels of PG I and/or PG I/PG II are closely related to AG and predict the risk of gastric cancer. We evaluated the performance of the chemiluminescent immunoassay-based Architect Pepsinogen I/II assay. Methods The evaluation consisted of determination of the precision, linearity, limit of blank (LoB), limit of detection (LoD) and method comparison with Eiken and Biohit assays. Results The total CVs were below 5% for both PG I and PG II. Acceptable linearity was observed for PG I and PG II in their respective reportable ranges. The PG I LoB was 0.317 ng/mL and the PG II LoB was 0.418 ng/mL, and LoDs were 0.412 ng/mL and 0.497 ng/mL, respectively. Correlation analysis indicated that results of the Architect assay were comparable to those of the Eiken and Biohit assays, but the three methods lead to different estimations of the cancer risk. Conclusion The overall analytical performance of Architect Pepsinogen I/II assay is acceptable for the detection of patients with suspected AG. The categorization results of gastric cancer risk showed some difference among test methods suggesting the need for harmonization among the methods from vendors.