Quantitative analysis of the immunosuppressant CP-690,550 in whole blood by column-switching high-performance liquid chromatography and mass spectrometry detection

2005 
A fast and accurate method to quantify the new immunosuppressive JAK3 inhibitor CP-690,550 in whole blood using a dual-pump liquid chromatography-liquid chromatography-mass spectromctry (LC/LC-MS) system was developed andvalidated in nonhuman primate blood. Before injection, blood samples were prepared by precipitation with a reagent that included methanol and acetonitrile (30:70, vol/vol) along with the internal standard (C'P-istd). Column-switching LC/LC'-MS analysis used online extraction followed by separation on a C 8 analytic column and MS detection of the [M + H] + CP-690,550 (m/z = 313.1) and CP internal standard (m/z = 288.1). Linearity was always better than r 2 = 0.99 (n = 7) for CP-690,550 (range 2.5-750 ng/mL), with a lower limit of quantification (LLOQ) of 2.5 ng/mL. The intrarun accuracy and precision ranged from 103.0% to 105.4% and 2.7% to 4.3%, respectively (n = 5), and the interday precision ranged from 8.7% to 11.1%, and the interday accuracy ranged from 98.1% to 103.8% of nominal values (n = 14). The injection repeatability for the method was 1.3% (n = 7). Except for the LLOQ, the intraday accuracy and precision in human blood were also within 15% (n = 5). The combination of simple sample preparation and short analytic run time of this sensitive procedure makes it effective for monitoring the concentration of CP-690,550 in whole blood in organ-transplant recipients.