Low Dose Rate Tests of Plastic Scintillator Plates and Wavelength Shifting Fibers

1994 
A low dose rate evaluation of several scintillators and wavelength shifters of interest to the CEBAF program was performed with a {sup 60}Co line source to a total dose of 160 Gray at a rate of 0.5 Gy/hr in a slowly flowing oxygen atmosphere. The samples included the following: (1) a 2 meter piece of Bicron BC412 scintillator being extensively used in the CEBAF Large Acceptance Spectrometer, (2) a 1 meter long piece of green scintillator doped with a uniquely fast green fluor (G2), (3) wavelength shifting fiber doped also with G2, and (4) grooved acrylic-based scintillator that is readout with the WLS fiber. The data indicates that the BC412 shows no loss in intrinsic light output, but does have an attenuation loss amounting to 10% at 1 meter and 16% at 2 meters. The fast green scintillator (and fibers) showed no damage. When read out by WLS fibers, the acrylic scintillator displayed a loss in intrinsic light output, but no change in attenuation.