Simultaneous cavity-enhanced and cavity ringdown absorption spectroscopy using optical feedback

2014 
We present a scheme of optical feedback cavity-enhanced absorption spectroscopy (OF-CEAS) including a fast optical switch to produce cavity ringdown spectra (OF-CRDS) simultaneously. This also works as a dynamically adjustable variable attenuator allowing to compensate for reduced signal levels in correspondence with absorption lines. For this, an acousto-optic deflector is used in a double-pass configuration to eliminate the single-pass frequency shift, which is incompatible with optical feedback. This is probably the most effective device providing the required fast response and the high extinction ratio necessary to perform clean ringdown measurements. The resulting direct comparison of OF-CEAS and OF-CRDS shows that these produce almost equivalent spectral data, with 0.3 % maximal difference at the top of an absorption line having a signal-to-noise ratio (S/N) of 3,300. OF-CEAS is largely winning on the short-term noise level while OF-CRDS appears to be more immune from interference fringes, delivering cleaner spectra after longer averaging.