Ambulatory blood pressure monitoring : validation of the accuracy and reliability of the TM-2420 according to the AAMI recommendations

1991 
The accuracy of the TM-2420 ambulatory blood pressure monitor (AD these differences were not significant (NS). The monitor was also assessed against direct intra-brachial artery pressure in 12 subjects (36 readings). The mean difference between the monitor and simultaneous individual intra-arterial reading was -9.5 mmHg for SBP (P less than 0.001) and 3.7 mmHg for DBP (P less than 0.001). The SDD's between methods were 12 mmHg for SBP and 5.0 mmHg for DBP. Use of the monitor in general clinical practice in 100 patients was also assessed. The rate of errors was low (6.8%), and the device found to be acceptable.(ABSTRACT TRUNCATED AT 250 WORDS)