Evaluation of a calibration transfer between a benchtop and portable mid-InfraRed spectrometer for cocaine classification and quantification
2019
Abstract A portable Fourier Transform mid-InfraRed (FT-MIR) spectrometer using Attenuated Total Reflectance (ATR) sampling is used for daily routine screening of seized powders. Earlier, ATR-FT-MIR combined with Support Vector Machines (SVM) algorithms resulted in a significant improvement of the screening method to a reliable and straightforward classification and quantification tool for both cocaine and levamisole. However, can this tool be transferred to new (hand-held) devices, without loss of the extensive dataset?The objective of this study was to perform a calibration transfer between a newly purchased bench-top (BT) spectrometer and a portable (P) spectrometer with existing calibration models. Both instruments are from the same brand and have identical characteristics and acquisition parameters (FT instrument, resolution of 4 cm−1 and wavenumber range 4000 to 500 cm−1). The original SVM classification model (n = 515) and SVM quantification model (n = 378) were considered for the transfer trial. Three calibration transfer strategies were assessed: 1) adjustment of slope and bias; 2) correction of spectra from the new instrument BT to P using piecewise direct standardization (PDS) and 3) building a new mixed instrument model with spectra of both instruments. For each approach, additional cocaine powders were measured (n = 682) and the results were compared with GC-MS and GC-FID. The development of a mixed instrument model was the most successful in terms of performance. The future strategy of a mixed model allows applying the models, developed in the laboratory, to portable instruments that are used on-site, and vice versa. The approach offers opportunities to exchange data within a network of forensic laboratories using other FT-MIR spectrometers.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
17
References
9
Citations
NaN
KQI