LC-MS/MS METHOD DEVELOPMENT AND VALIDATION FOR THE DETERMINATION OF CARDIAZOL IN HUMAN PLASMA

2019 
The total chromatographic run time was 2.5 min and the elution of cardiazol and IS (difenoconazole) occurred at ~2.15 and 1.98 min, respectively. A linear response function was established at 1-100 ng/ml for cardiazol and difenoconazole in human plasma. The % mean recovery for cardiazol in LQC, MQC and HQC was 102.8 %, 100.3 % and 95.9 %. The lowest concentration with the RSD<20% was taken as LLOQ and was found to be 1.10 ng/ml for cardiazol. The % accuracy of LLOQ samples prepared with the different biological matrix lots was found 109.7 %, which were found within the range of 80.00-120.00 % for the seven different plasma lots. % CV for LLOQ samples was observed as 11.9 %, which are within 20.00% of the acceptance criteria. The within-run coefficients of variation ranged between 0.311 % and 0.601 % for cardiazol. The within-run percentages of nominal concentrations ranged between 99.80 % and 100.41 % for cardiazol. The between-run coefficients of variation ranged between 0.332 % and 0.615 % for cardiazol. The between-run percentages of nominal concentrations ranged between 98.18 % and 101.21 % for cardiazol.