Biomechanical Behavior of an Hydroxyapatite-Coated Traditional Hip Stem and a Short One of Similar Design: Comparative Study Using Finite Element Analysis

2021 
Abstract Background The objective is to compare, by the means of finite elements analysis, the biomechanical behavior of a conventional stem of proven performance with a short stem based on the same fixation principles. Methods A 3D femur was modeled from CT scan data, and real bone density measures were incorporated into it. Load stresses were applied to that bone in 3 different scenarios: without prosthesis, with the conventional stem, and with the short stem. Different bone loading patterns were compared by Gruen’s zones both visually and statistically using Welch’s test. Results The implantation of a stem generates a certain degree of stress shielding in the surrounding bone, but the pattern of the change is very similar in the compared stem models. Although there is statistical significance (P Conclusions The bone loading patterns of the traditional stem and the short stem are very similar. Although there is no evidence of a link between biomechanics and clinical outcomes, our results may suggest that theoretical advantages of short stems can be exploited without the fear of altering bone loading patterns.