Explant rates of electrical neuromodulation devices in 1177 patients in a single center over an 11-year period.

2020 
Introduction The publication of explant rates has established risk factors and a definitive objective outcome of failure for spinal cord stimulation (SCS) treating neuropathic pain. We present a UK study analyzing explants of electrical neuromodulation devices for different conditions over 11 years in a single center specializing in neuromodulation. Methods A retrospective analysis was performed using a departmental database between 2008 and 2019. Explants were analyzed according to condition, mode of stimulation and other demographics using logistic regression and Kaplan-Meier graphs with log-rank (Mantel-Cox) test. Results Out of a total of 1177 patients, the explant rate was 17.8% at 5 years and 25.2% at 10 years. Loss of efficacy was the most frequent reason for explant 119/181 (65%). Multivariant regression analysis indicated patients with back pain without prior surgery had a reduced risk of explant (p=0.03). Patients with SCS systems that had 10 kHz, options of multiple waveforms, and rechargeable batteries also had a decreased risk of explant (p Conclusion These data contribute to a growing list of explant data in the scientific literature and give indications of what factors contribute to long-term utilization of electrical neuromodulation devices.