A fast position sensitive microstrip-gas-chamber detector at high count rate operation
2002
Testing of a newly developed position sensitive high count rate microstrip gas chamber (MSGC) detector at high count rate operation has been carried out at the Dutch–Belgian x-ray scattering beamline at the European Synchrotron Radiation Facility (Grenoble, France) with a high intensity x-ray beam. The measurements show local count rate capabilities up to ∼4.5×105 counts/s/channel. Experimental data taken with this detector are also shown. These tests show that both time resolution down to 1.5 ms/frame and a reliable operation at high counting rates can be achieved.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
15
References
14
Citations
NaN
KQI