Two Fibers Packed in Capillary Column for Gas Chromatographic Separation

2021 
The bare optical fiber of 0.12 mm o.d. has been introduced as the support material in gas chromatography. Two fused-silica fibers coated with polydimethylsiloxane (SE30) were longitudinally packed into a 2 m SE30 capillary column (SE30-CFs-packed SE30-CC), followed by studying chromatographic performances. With 0.012 g/mL SE30 solution coating onto the packed fibers, the efficiency per unit length of SE30-CFs-packed SE30-CC was 1695 plates/m (for the separation of n-hexadecane at 120°С). The retention power and sample loading capacity of this SE30-CFs-packed SE30-CC were improved over a single SE30-coated fused-silica fiber in a SE30-coated capillary annular column (SE30-CF/SE30-CC) as well as a conventional SE30 capillary column (HP-1). The results showed the contribution of the two coated fused-silica fibers to the retentivity and sample loading capacity of the column. With a large injection, SE30-CFs-packed SE30-CC provided better separation for hidden harmful solvents in beta-cypermethrin malathion than SE30-CF/SE30-CC and HP-1 of the same length.