Computerized Bone Densitometric Analysis: Operator-dependent Errors

1999 
PURPOSE: To determine the nature and relative frequency of operator-dependent data analysis errors in dual x-ray absorptiometry. MATERIALS AND METHODS: Over 40 months, 2,528 dual x-ray absorptiometric examinations of the forearm, femoral neck, and lumbar spine were performed by 11 technologists by using standard techniques and software. Each analysis was reviewed by a radiologist; errors were recorded and corrected. RESULTS: There were no forearm analysis errors. There were 24 (0.9%) femoral neck analysis errors, of which 23 resulted from misplacement of the analysis region. There were 33 (1.3%) spinal analysis errors, of which 24 resulted from misplacement of intervertebral disk space markers. Analysis errors of the femur and spine resulted in six misdiagnoses (0.2%). CONCLUSION: Misdiagnosis due to analysis errors is rare. Femoral neck analysis errors were easily detectable, but accurate spinal analyses depended on accurate identification of vertebral end plates and posterior elements. Nonetheless, thes...