Beamline Spectroscopy of Integrated Circuits With Hard X-ray Transition Edge Sensors at the Advanced Photon Source

2020 
At Argonne National Laboratory, we are developing hard X-ray (2 to 20 keV) Transition Edge Sensor (TES) arrays for beamline science. The significantly improved energy resolution provided by superconducting detectors compared to semiconductor-based energy-dispersive detectors, but with better collection efficiency than wavelength-dispersive instruments, will enable greatly improved X-ray emission and absorption spectroscopic measurements. A prototype instrument with 24 microwave-frequency multiplexed pixels is now in testing at the Advanced Photon Source (APS) 1-BM beamline. Initial measurements show an energy resolution ten times better than the silicon-drift detectors currently available to APS beamline users, and in particular demonstrate the ability to resolve closely-spaced emission lines in samples containing multiple transition metal elements, such as integrated circuits. Comparing fluorescence spectra of integrated circuits measured with our TESs at the beamline to those measured with silicon detectors, we find emission lines and elements largely hidden from a semiconductor-based detector but well resolved by a TES. This directly shows the strengths of TES-based instruments in fluorescence mapping.