Liquid-phase chemical and biochemical detection using fully integrated magnetically actuated complementary metal oxide semiconductor resonant cantilever sensor systems

2007 
A novel resonant cantilever sensor system for liquid-phase applications is presented. The monolithic system consists of an array of four electromagnetically actuated cantilevers with transistor-based readout, an analog feedback circuit, and a digital interface. The biochemical sensor chip with a size of 3 mm × 4.5 mm is fabricated in an industrial complementary metal oxide semiconductor (CMOS) process with subsequent CMOS-compatible micromachining. A package, which protects the electrical components and the associated circuitry against liquid exposure, allows for a stable operation of the resonant cantilevers in liquid environments. The device is operated at the fundamental cantilever resonance frequency of ∼200 kHz in water with a frequency stability better than 3 Hz. The use of the integrated CMOS resonant cantilever system as a chemical sensor for the detection of volatile organic compounds in liquid environments is demonstrated. Low concentrations of toluene, xylenes, and ethylbenzene in deionized wat...