Accuracy of an oscillometric blood pressure monitor in anesthetized pigs

2018 
The purpose of this study is to evaluate the accuracy of an oscillometric blood pressure monitor in anesthetized pigs. Invasive blood pressure (IBP) and noninvasive blood pressure (NIBP) measurements were taken using a DRE Waveline Pro multiparameter monitor at four different time points in 17 pigs undergoing injectable anesthesia. NIBP measurements were taken on both the thoracic and pelvic limbs. Bland Altman analysis was used to assess agreement between methods and a linear mixed-effects model was used to evaluate the effect of cuff position and blood pressure on bias. Invasive systolic arterial pressure (SAP) ranged between 112 and 161 mmHg (mean ± SD: 138.8 ± 13.3; median: 139.5). Invasive diastolic arterial pressure (DAP) ranged between 60 and 104 mmHg (mean ± SD: 86.0 ± 9.1; median: 87.0). Invasive mean arterial pressure (MAP) ranged between 79 and 121 mmHg (mean ± SD: 103.2 ± 9.3; median 103.0). Only the diastolic and mean measurements obtained from the pelvic limb met criteria outlined by the Ame...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    0
    Citations
    NaN
    KQI
    []