Measurement of absolute detection efficiencies of a microchannel plate using the charge transfer reaction

2011 
A new method for determining the absolute detection efficiencies of a microchannel plate (MCP) detector is presented. This method uses a charge transfer reaction and can be applied to determine not only the detection efficiencies for ions but also those for neutral atoms. The measurements were carried out for Ar+ and neutral rare gas atoms (Ne, Ar, Kr and Xe) in the 0.5?4.8?keV energy range. The detection efficiencies for all species increase with increasing impact energy and approach the open area ratio of the MCP used (about 50%). The measured detection efficiencies were found to scale with the Lindhard, Scharff and Schiott formula for electronic stopping power at keV energies.